Fluke 9170-E-R Field Metrology Well –45°C to 140°C

The Fluke Calibration 9170-E-R Metrology Well Insert “E” is made of aluminum and features miscellaneous metric holes, along with a 0.25-inch reference hole, providing versatility in temperature calibration applications. With a temperature range spanning from -45°C to 140°C, this insert ensures accurate and reliable calibration across a wide range of temperatures.

SKU:79314

MME Shopping Event

Hurry and get discounts on Most of MME Products up to 20%

Call for Price

12People watching this product now!

Fluke Calibration 9170-E-R Metrology Well Insert “E”, Al, Misc Metric Holes, w/ 0.25-inch Ref Hole –45°C to 140°C

The Fluke 9170-E-R Metrology Well Insert “E” is made of aluminum and features miscellaneous metric holes, along with a 0.25-inch reference hole, providing versatility in temperature calibration applications. With a temperature range spanning from -45°C to 140°C, this insert ensures accurate and reliable calibration across a wide range of temperatures.

Crafted with durable aluminum material, the 9170-E-R insert delivers consistent performance and reliability. Its miscellaneous metric holes are designed to accommodate various calibration requirements, while the 0.25-inch reference hole serves as a standard for calibration comparisons.

Whether utilized in laboratories, manufacturing facilities, or field environments, the Fluke Calibration 9170-E-R Metrology Well Insert “E” serves as an essential tool for metrology professionals and technicians, facilitating precise temperature calibration with confidence.

fluke 9170 series

The Model 9170 stands out in the series by attaining the lowest temperatures, reaching as low as -45°C even under normal room conditions. With a stability rating of ±0.005°C across its entire temperature range (up to 140°C) and an immersion depth of 160 mm (6.3 in), it offers remarkable precision. Featuring axial uniformity of ±0.02°C and radial uniformity of ±0.01°C, this model boasts outstanding uncertainty budgets, making it an ideal choice for a wide range of pharmaceutical and other applications.

Fluke 9170 ER Specifications:

Specifications
Range (at 23°C ambient)–45°C to 140°C (–49°F to 284°F)
Display accuracy±0.1°C full range
Stability±0.005°C full range
Axial uniformity±0.1°C at –45°C
±0.04°C at –35°C
±0.02°C at 0°C
±0.07°C at 140°C
Radial uniformity±0.01°C full range
Loading effect (with a 6.35 mm reference probe and three 6.35 mm probes)±0.02°C at –45°C
±0.005°C at –35°C
±0.01°C at 140°C
Hysteresis0.025°C
Well depth160 mm (6.3 in)
Resolution0.001°C
DisplayLCD, °C or °F, user-selectable
Key padTen key with decimal and ± button. Function keys, menu key, and °C / °F key.
Cooling time44 min: 23°C to –45°C
19 min: 23°C to –30°C
19 min: 140°C to 23°C
Heating time32 min: 23°C to 140°C
45 min: –45°C to 140°C
Size (H x W x D)366 x 203 x 323 mm (14.4 x 8 x 12.7 in)
Weight14.2 kg (31.5 lb)
Power115 V AC (±10%), or
230 V AC (±10%), 50/60 Hz, 550 W
Computer interfaceRS-232 Interface with 9930 Interface-it control software included
Traceable calibration (NIST)Data at –45°C, 0°C, 50°C, 100°C, and 140°C
SpecificationsBuilt-in Reference Input
Temperature range–200°C to 962°C (–328°F to 1764°F)
Resistance range0 Ω to 400 Ω, auto-ranging
CharacterizationsITS-90 subranges 4, 6, 7, 8, 9, 10, and 11 Callendar-Van Dusen (CVD): R0, a, b, d
Resistance accuracy0 Ω to 20 Ω: 0.0005 W
20 Ω to 400 Ω: 25 ppm
Temperature accuracy
(does not include probe uncertainty)
10 Ω PRTs:
±0.013°C at 0°C
±0.014°C at 155°C
±0.019°C at 425°C
±0.028°C at 700°C
25 Ω and 100 Ω PRTs:
±0.005°C at –100°C
±0.007°C at 0°C
±0.011°C at 155°C
±0.013°C at 225°C
±0.019°C at 425°C
±0.027°C at 661°C
Resistance resolution0 Ω to 20 Ω: 0.0001 Ω
20 Ω to 400 Ω: 0.001 Ω
Measurement period1 second
Probe connection4-wire with shield, 5-pin DIN connector
CalibrationNVLAP accredited (built-in reference input only), NIST-traceable calibration provided
0 reviews
0
0
0
0
0

There are no reviews yet.

Be the first to review “Fluke 9170-E-R Field Metrology Well –45°C to 140°C”

Brand

Fluke

Fluke Corporation

Since its founding in 1948, Fluke has helped define and grow a unique technology market, providing testing and troubleshooting capabilities that have grown to mission critical status in manufacturing and service industries. Every new manufacturing plant, office, hospital, or facility built today represents another potential customer for Fluke products. From industrial electronic installation, maintenance and service, to precision measurement and quality control, Fluke tools help keep business and industry around the globe up and running. Typical customers and users include technicians, engineers, metrologists, medical-device manufacturers, and computer network professionals — people who stake their reputations on their tools, and use tools to help extend their personal power and abilities.